share:
delim
delim
Pesticides and Brominated Flame Retardants by EPA Method 527 on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Pesticides and Brominated Flame Retardants by EPA Method 527 on Rxi-5Sil MS

Peaks
1.Atrazine
2.Vinclozoline
3.Malathion
4.Chlorpyrifos
5.Terbufos sulfone
6.Nitrofen
7.Kepone
8.Norflurazon
Peaks
9.Triphenyl phosphate
10.Bifenthrin
11.BDE-47
12.Mirex
13.BDE-100
14.BDE-99
15.Fenvalerate
16.Esfenvalerate
17.BDE-153
image
GC_EV1215
ColumnRxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620)
using IP Deactivated Guard Column 5 m, 0.53 mm ID (cat.# 10045)
with Universal Press-Tight Connectors (cat.# 20429)
SamplePBDE Mix (cat.# 33098)
Pesticides Mix #1, Method 527 (cat.# 33007)
Pesticides Mix #2, Method 527 (cat.# 33008)
Internal Standard, Method 527 (cat.# 33010)
Surrogate Standard, Method 527 (cat.# 33009)
Diluent:ethyl acetate:methylene chloride (1:1)
Conc.:80 pg/µL (1 ng on-column)
Injection
Inj. Vol.:12.5 µL splitless (hold 0.583 min)
Liner:Gooseneck Splitless (4 mm) w/Semivolatiles Wool (cat.# 20799-231.5)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:40 °C (hold 0.60 min) to 320 °C at 30 °C/min (hold 1.07 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:-70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:3.33 min
Acquisition Range:45 - 750 amu
Spectral Acquisition Rate:10 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesCarrier Gas Flow: 2 mL/min. corrected constant flow via pressure ramps
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

BFR

,

527

,

EPA Method 527

,

persistent organic pollutants

,

TOF

,

TOFMS

,

halogenated pesticide

,

pesticides

,

POP

,

brominated flame retardants

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us