Pure Chromatography
www.restek.com
www.restek.com



This chromatogram may contain analytes that are synonyms of your search terms.
GCxGC Contour Plot of Riser Pipe Oil in 2010 Gulf Oil Spill (Rxi-17Sil MS and Rxi-1ms)

GC_EV1229
Column | Rxi-17Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123) |
Rxi-1ms 1.2 m, 0.15 mm ID, 0.15 µm (cat.# 43800) | |
Sample | |
Diluent: | Methylene chloride |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Premium 4mm Single Taper w/Wool (cat.# 23303.1) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | Rxi-17Sil MS: 40 °C (hold 1 min) to 320 °C at 3.5 °C/min (hold 4 min) Rxi-1ms: 45 °C (hold 1 min) to 325 °C at 3.5 °C/min (hold 4 min) |
Carrier Gas | He, corrected constant flow (2.2 mL/min) |
Modulation | |
Modulator Temp. Offset: | 20 °C |
Second Dimension Separation Time: | 4 sec |
Hot Pulse Time: | 1.4 sec |
Cool Time between Stages: | 0.6 sec |
Detector | TOFMS |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | TOF |
Source Temp.: | 250 °C |
Electron Energy: | 70 eV |
Mass Defect: | 50 mu/100 u |
Solvent Delay Time: | 5 min |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Acquisition Range: | 45-550 amu |
Spectral Acquisition Rate: | 200 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | Rxi-1ms (cat.# 43800) is a 10 m column. A 1.2 m section was used as the second dimension column. |
Acknowledgement | Dr. Ed Overton, Professor Emeritus at Louisiana State University, for providing the riser pipe oil sample. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.