share:
delim
delim
Semivolatile Organics in Drinking Water on Rxi-5Sil MS by EPA Method 525.2

Semivolatile Organics in Drinking Water on Rxi-5Sil MS by EPA Method 525.2

Peaks
1.Isophorone
2.2-nitro-m-xylene (SS)
3.Naphthalene
4.Dichlorvos (DDVP)
5.Hexachlorocyclopentadiene
6.EPTC
7.Mevinphos
8.Butylate
9.Vernolate
10.Dimethyl phthalate
11.Etridiazole
12.Pebulate
13.2,6-Dinitrotoluene
14.Acenaphthylene
15.Acenaphthene-d10 (IS)
16.Chlorneb
17.Tebuthiuron
18.2,4-Dinitrotoluene
19.Molinate
20.Diethyl phthalate
21.Fluorene
22.Propachlor
23.Ethoprop (ethoprophos)
24.Cycloate
25.Trifluralin
26.Chlorpropham
27.Atraton
28.Hexachlorobenzene
29.Prometon
30.Simazine
31.Atrazine
32.Propazine
33.Pentachlorophenol
34.Terbufos
35.Diazinon
36.Propyzamide
37.Disulfoton
38.Methylparaoxon
39.Chlorothalonil
40.Phenanthrene-d10 (IS)
41.Terbacil
42.Phenanthrene
43.Anthracene
44.Metribuzin
45.Alachlor
46.Simetryn
47.Ametryn
Peaks
48.Prometryn
49.Terbutryn
50.Dibutyl phthalate
51.Bromacil
52.Metolachlor
53.Chlorpyrifos
54.Cyanazine (Bladex)
55.DCPA methyl ester (Chlorthal-dimethyl)
56.Triadimefon
57.Diphenamid
58.MGK-264 (isomer A)
59.MGK-264 (isomer B)
60.Merphos
61.Heptachlor epoxide
62.Fluoranthene
63.Stirofos
64.Butachlor
65.Disulfoton sulfone
66.Fenamiphos
67.Napropamide (Devrinol®)
68.Pyrene-d10 (SS)
69.trans-Nonachlor
70.Pyrene
71.Tribufos (merphos oxide)
72.Tricyclazole (Beam)
73.Carboxin
74.Chlorbenzilate
75.Norflurazon
76.Benzyl butyl phthalate
77.Bis(2-ethylhexyl) adipate
78.Hexazinon (Velpar®)
79.Triphenylphosphate (SS)
80.Benzo[a]anthracene
81.Bis(2-ethylhexyl)phthalate
82.Chrysene-d12 (IS)
83.Chrysene
84.Fenarimol
85.cis-Permethrin
86.Di-n-octyl phthalate
87.trans-Permethrin
88.Benzo[b]fluoranthene
89.Benzo[k]fluoranthene
90.Fluridone (Sonar®)
91.Benzo[a]pyrene
92.Perylene-d12 (SS)
93.Indeno[1,2,3-cd]pyrene
94.Dibenzo[a,h]anthracene
95.Benzo[ghi]perylene
image
GC_EV1247
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMethod 525.2 internal standard mix (cat.# 31825)
Method 525.2 surrogate standard mix (cat.# 31826)
Method 525.2 semivolatile mix (revised) (cat.# 31899)
Organochlorine pesticide mix #2 (rev), Method 525.2 (cat.# 33011)
Organonitrogen pesticide mix #1 (rev), Method 525.2 (cat.# 33012)
Organophosphorus pesticide mix #1 (rev), Method 525.2 (cat.# 33013)
Method 525.2 nitrogen/phosphorus pesticide mix #2 (cat.# 32423)
Diluent:Acetone
Conc.:8 µg/mL (IS/SS 25 μg/mL)
Injection
Inj. Vol.:1 µL pulsed splitless (hold 0.59 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:270 °C
Pulse Pressure:30 psi (206.8kPa)
Pulse Time:0.64 min
Purge Flow:100 mL/min
Oven
Oven Temp.:50 °C (hold 1 min) to 285 °C at 30 °C/min to 305 °C at 3 °C/min to 320 °C at 30 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
Linear Velocity:39.723 cm/sec @ 40 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:270 °C
Quad Temp.:150 °C
Solvent Delay Time:4 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:35-550 amu
Scan Rate:5.36 scans/sec
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 525

,

sky liner

,

Sky single taper

,

liner with wool

,

525

,

525.2

,

drinking water

,

semivolatiles

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us