EPA 522 Analyte Mix on Rtx-624 w/5 m Integra-Guard (30 m x 0.25 mm x 1.4 μm)

EPA 522 Analyte Mix on Rtx-624 w/5 m Integra-Guard (30 m x 0.25 mm x 1.4 μm)

PeakstR (min)
1.Tetrahydrofuran-D85.59
2.1,4-Dioxane-d86.51
3.1,4-Dioxane6.55
image
GC_EV1250
ColumnRtx-624 w/5 m Integra-Guard, 30 m, 0.25 mm ID, 1.40 µm (cat.# 10968-124)
SampleTetrahydrofuran-d8 (cat.# 30112)
1,4-Dioxane-d8 (cat.# 30614)
1,4-Dioxane (cat.# 30287)
Diluent:Dichloromethane
Conc.:50 ng/mL
Injection
Inj. Vol.:10 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:120 °C
Purge Flow:80 mL/min
Oven
Oven Temp.:35 °C (hold 1 min) to 120 °C at 12 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
Linear Velocity:30.556 cm/sec @ 35 °C
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
15.246, 78, 80 50
26.396, 88, 64, 62, 58 75
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:5.2 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
NotesLarge volume splitless injection for EPA Method 522 without a PTV and without a press-fit connector for the retention gap.
AcknowledgementM. Misselwitz, J. Cochran
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

dioxane

,

4-dioxane

,

large volume splitless injection

,

EPA 522

,

drinking water

,

1

,

LVSI

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us