Semivolatile GC-MS Tune Verification Mix on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Semivolatile GC-MS Tune Verification Mix on Rxi-5Sil MS

Peaks
1.Pentachlorophenol
2.Decafluorotriphenylphosphine
Peaks
3.Benzidine
4.p,p'-DDT
image
GC_EV1273
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleGC-MS tuning mixture (cat.# 31615)
Diluent:Dichloromethane
Conc.:50 ng/µL
Injection
Inj. Vol.:1 µL pulsed splitless (hold 0.59 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:270 °C
Pulse Pressure:30 psi (206.8kPa)
Pulse Time:0.64 min
Purge Flow:100 mL/min
Oven
Oven Temp.:40 °C (hold 1 min) to 280 °C at 25 °C/min to 320 °C at 5 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:276 °C
Quad Temp.:150 °C
Solvent Delay Time:2.19 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:35-550 amu
Scan Rate:5.36 scans/sec
InstrumentAgilent 7890A GC & 5975C MSD
Notes7890 Siltek-treated EZ Twist Top split/splitless injection port (cat.# 22178)
Flip Seal dual Vespel ring inlet seal (cat.# 23411)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Appendix IX

,

SVOA

,

8270

,

semivolatiles

,

5sil MS

,

RXI

,

rxi-5sil

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us