Nitrosamines in Drinking Water on Rxi-624Sil MS by Modified EPA Method 521 Utilizing CSR-LVSI
This chromatogram may contain analytes that are synonyms of your search terms.

Nitrosamines in Drinking Water on Rxi-624Sil MS by Modified EPA Method 521 Utilizing CSR-LVSI

Peaks
1.N-Nitrosodimethylamine-d6
2.N-Nitrosodimethylamine
3.N-Nitrosomethylethylamine
4.N-Nitrosodiethylamine
Peaks
5.N-Nitrosodipropylamine-d14
6.N-Nitrosodipropylamine
7.N-Nitrosopyrrolidine
8.N-Nitrosopiperidine
9.N-Nitrosodibutylamine
image
GC_EV1286
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029)
with universal Press-Tight connectors (cat.# 20429)
SampleN-Nitrosodimethylamine-d6 (cat.# 33910)
N-Nitrosodi-n-propylamine-d14 (cat.# 33911)
Nitrosamine calibration mix, method 521 (cat.# 31898)
Diluent:Dichloromethane
Conc.:10 ng/mL
Injection
Inj. Vol.:10 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Purge Flow:80 mL/min
Oven
Oven Temp.:38 °C (hold 1 min) to 160 °C at 12 °C/min to 200 °C at 5 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorAgilent 5975C
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
16.980, 74, 46, 43, 42 23
27.988, 56, 43, 42 30
38.9102, 56, 44, 42 30
410130, 78, 70, 58, 46, 43, 42 15
512.05100, 68, 42, 41 30
614.5158, 116, 84, 57, 41 23
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:6.9 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us