share:
delim
delim
50 µL CSR-LVSI with 5 ng Nitrosamines, THF, and 1,4-Dioxane On-Column
This chromatogram may contain analytes that are synonyms of your search terms.

50 µL CSR-LVSI with 5 ng Nitrosamines, THF, and 1,4-Dioxane On-Column

Peaks
1.Tetrahydrofuran
2.1,4-Dioxane
3.N-Nitrosodimethylamine
4.N-Nitrosomethylethylamine
Peaks
5.N-Nitrosodiethylamine
6.N-Nitrosopyrrolidine
7.N-Nitrosodi-n-propylamine
8.N-Nitrosopiperidine
9.N-Nitrosodi-n-butylamine
* Toluene contaminant
image
GC_EV1342
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 13653)
using Rxi guard column 10 m, 0.53 mm ID (cat.# 10073)
with BGB P/N: 2553LD
Sample1,4-Dioxane (cat.# 30287)
Nitrosamine calibration mix, Method 521 (cat.# 31898)
Tetrahydrofuran (THF) (cat.# 30414)
Diluent:Dichloromethane
Conc.:0.1 µg/mL
Injection
Inj. Vol.:50 µL splitless (hold 1.5 min)
Liner:Custom Premium single taper with 15 mg quartz wool
Inj. Temp.:275 °C
Purge Flow:100 mL/min
Oven
Oven Temp.:35 °C (hold 1.5 min) to 50 °C at 50 °C/min (hold 7.1 min) to 320 °C at 11.12 °C/min (hold 1.5 min)
Carrier GasHe, constant flow
Flow Rate:5.08 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
19.7042, 46, 71, 72, 78, 80 40
210.2558, 62, 64, 88, 96 40
310.7542, 43, 46, 48, 74, 80 40
411.5042, 44, 56, 88, 102 40
514.0056, 68, 70, 78, 86, 100, 130, 144 25
616.2055, 56, 84, 114 40
717.5099, 116, 141, 158 40
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:9.7 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us