0.50 ng/L N-Nitrosodimethylamine (NDMA) Laboratory Fortified Drinking Water Sample Extract on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

0.50 ng/L N-Nitrosodimethylamine (NDMA) Laboratory Fortified Drinking Water Sample Extract on Rxi-5Sil MS

PeaksConc.
(ng/mL)
m/z
1.N-Nitrosodimethylamine-d61080
PeaksConc.
(ng/mL)
m/z
2.N-Nitrosodimethylamine0.02574
Sample fortified at 0.5 ng/L N-nitrosodimethylamine. Final extract concentration is 0.025 ng/mL.
image
GC_EV1348
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 13653)
using Rxi guard column 10 m, 0.53 mm ID (cat.# 10073)
with SGE µ-union
Sample
Diluent:Dichloromethane
Injection
Inj. Vol.:50 µL splitless (hold 1.5 min)
Liner:Custom Premium single taper liner with 15 mg quartz wool
Inj. Temp.:275 °C
Purge Flow:100 mL/min
Oven
Oven Temp.:35 °C (hold 1.5 min) to 50 °C at 50 °C/min (hold 7.1 min) to 320 °C at 11.12 °C/min (hold 1.5 min)
Carrier GasHe, flow program
Flow Program:5.08 mL/min (hold 8.9 min) to 2.0 mL/min at 1.0 mL/min/min
DetectorMS
Mode:SIM
SIM Program:42, 43, 46, 48, 74, 80 m/z
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:9.9 min
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us