25 µg/L 8260 MegaMix Revised on Rtx-VMS
*Note: these compounds were not detected at 25 ppb when the sample was analyzed via unheated purge.

GC_EV1352
Column | Rtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915) | ||||||||||||
Sample | 8260B MegaMix calibration mix, revised (cat.# 30152) | ||||||||||||
8260 Internal standard mix (cat.# 30074) | |||||||||||||
8260 Surrogate mix (cat.# 30073) | |||||||||||||
Diluent: | Reverse-osmosis water | ||||||||||||
Conc.: | 25 µg/L (5 mL sample) | ||||||||||||
Injection | purge and trap split | ||||||||||||
Liner: | Premium 1 mm Split (cat.# 23333.1) | ||||||||||||
Purge and Trap | |||||||||||||
Instrument: | EST Encon | ||||||||||||
Trap Type: | Vocarb 3000 | ||||||||||||
Purge: | 11 min, flow 40 mL/min | ||||||||||||
Dry Purge: | 3 min, flow 40 mL/min | ||||||||||||
Desorb Preheat Temp.: | 255 °C | ||||||||||||
Desorb: | 1.0 min @ 260 °C, flow 30.9 mL/min | ||||||||||||
Bake: | 6 min @ 260 °C | ||||||||||||
Interface Connection: | injection port | ||||||||||||
Transfer Line Temp.: | 150 °C | ||||||||||||
Oven | |||||||||||||
Oven Temp.: | 45 °C (hold 4.5 min) to 100 °C at 12 °C/min to 240 °C at 25 °C/min (hold 1.32 min) | ||||||||||||
Carrier Gas | He, constant flow | ||||||||||||
Flow Rate: | 0.9 mL/min | ||||||||||||
Detector | MS | ||||||||||||
Mode: | Scan | ||||||||||||
Scan Program: |
| ||||||||||||
Transfer Line Temp.: | 230 °C | ||||||||||||
Analyzer Type: | Quadrupole | ||||||||||||
Source Type: | Inert | ||||||||||||
Drawout Plate: | 6mm ID | ||||||||||||
Source Temp.: | 230 °C | ||||||||||||
Quad Temp.: | 150 °C | ||||||||||||
Electron Energy: | 70 eV | ||||||||||||
Solvent Delay Time: | 1.5 min | ||||||||||||
Instrument | Agilent 7890A GC & 5975C MSD | ||||||||||||
Notes | Some purge-and-trap instruments show increased activity of dehydrohalogenate pentachloroethane during analysis, forming tetrachloroethene (also known as perchloroethylene, PCE). Symptoms of this problem are poor matching of PCE levels between sources or a PCE single analyte test. This revised 8260 MegaMix standard is identical to the original, minus pentachloroethane, and was made for testing that does not required reporting of pentachloroethane. |