Diesel #2/Mineral Oil Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)

Diesel #2/Mineral Oil Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)

PeakstR (sec)
1.C12203.5
2.C16349.2
3.C20468.8
image
GC_EV1378
ColumnRxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905)
SampleDiesel #2/mineral oil (cat.# 31676)
Diluent:Hexane
Conc.:5,000 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:17.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc).
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Mineral Oil

,

Hump-o-gram

,

Total Petroleum Hydrocarbons

,

Total Recoverable Petroleum Hydrocarbons

,

Alkanes

,

Diesel

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us