Diesel #2/Motor Oil Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)
Peaks | tR (sec) | |
---|---|---|
1. | C16![]() | 349.1 |
2. | C18![]() | 412.1 |
3. | C22![]() | 521.2 |
4. | C24![]() | 569.3 |

GC_EV1379
Column | Rxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905) |
Sample | Diesel #2/motor oil (cat.# 31682) |
Diluent: | Hexane |
Conc.: | 5,000 ng/µL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 10:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 17.5 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.75 mL/min |
Detector | FID @ 420 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Actual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc). |