Pure Chromatography
www.restek.com
www.restek.com



This chromatogram may contain analytes that are synonyms of your search terms.
Used Motor Oil Composite Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)
Peaks | tR (sec) | |
---|---|---|
1. | C40![]() | 856.7 |

GC_EV1382
Column | Rxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905) |
Sample | Used motor oil composite standard (cat.# 31465) |
Diluent: | Methylene chloride |
Conc.: | 50,000 ng/µL |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 275 °C |
Split Vent Flow Rate: | 87.5 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.75 mL/min |
Detector | FID @ 420 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
Notes | Actual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc). |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Total Petroleum Hydrocarbons
,Total Recoverable Petroleum Hydrocarbons
,Alkanes
,Diesel
,Hump-o-gram
,Motor Oil
,Used Motor Oil
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.