share:
delim
delim
EPA Method 8061A Phthalates on Rtx-440

EPA Method 8061A Phthalates on Rtx-440

Peaks
1.Dimethyl phthalate
2.Diethyl phthalate
3.Benzyl benzoate (IS)
4.Diisobutyl phthalate
5.Di-n-butyl phthalate
6.Bis(2-methoxyethyl) phthalate
7.Bis[4-methyl-2-pentyl] phthalate isomer 1
8.Bis[4-methyl-2-pentyl] phthalate isomer 2
Peaks
9.Bis(2-ethoxyethyl) phthalate
10.Dipentyl phthalate
11.Di-n-hexyl phthalate
12.Benzyl butyl phthalate
13.Bis(2-n-butoxyethyl) phthalate
14.Bis(2-ethylhexyl) phthalate
15.Dicyclohexyl phthalate
16.Di-n-octyl phthalate
17.Dinonyl phthalate
image
GC_EV1406
ColumnRtx-440, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12923)
SampleEPA Method 8061A phthalate esters mixture (cat.# 33227)
Benzyl benzoate (cat.# 31847)
Diluent:Methylene chloride
Conc.:80 µg/mL EPA 8061A phthalates (20 μg/mL internal standard)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 3.5 mm Precision liner w/wool (cat.# 23320.1)
Inj. Temp.:280 °C
Oven
Oven Temp.:150 °C (hold 1 min) to 300 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant linear velocity
Linear Velocity:50 cm/sec, 22.9 psi, 157.9 kPa @ 150 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1250-500-
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Electron Energy:70 eV
Solvent Delay Time:2 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesThe linear velocity is 50 cm/sec, which is approximately 1.75 mL/min at 150 °C.
The MS scan interval is 0.1 sec.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Phthalates

,

GC-MS

,

Impurities

,

Rtx-440

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us