Phthalates (50 µg/mL) on Rtx-440 (Extended List, Full Scan Mode)
This chromatogram may contain analytes that are synonyms of your search terms.

Phthalates (50 µg/mL) on Rtx-440 (Extended List, Full Scan Mode)

Peaks
1.Dimethyl phthalate
2.Dimethyl isophthalate
3.Diethyl phthalate
4.Benzyl benzoate
5.Diisobutyl phthalate
6.Di-n-butyl phthalate
7.Bis(2-methoxyethyl) phthalate
8.Bis[4-methyl-2-pentyl] phthalate isomer 1
9.Bis[4-methyl-2-pentyl] phthalate isomer 2
10.Bis(2-ethoxyethyl) phthalate
11.Di-n-pentyl phthalate
12.Butyl cyclohexyl phthalate
13.Butyl 2-ethylhexyl phthalate
14.Di-n-hexyl phthalate
15.Butyl octyl phthalate
16.Butyl benzyl phthalate
17.Hexyl-2-ethylhexyl phthalate
18.Butyl isodecyl phthalate
19.Bis(2-ethylhexyl)hexahydro phthalate
20.Bis(2-n-butoxyethyl) phthalate
Peaks
21.Dicyclohexyl phthalate
22.Bis(2-ethylhexyl) phthalate
23.Butyl-n-decyl phthalate
24.Diphenyl phthalate
25.Bis(4-methylcyclohexyl) phthalate isomer 1
26.Bis(4-methylcyclohexyl) phthalate isomer 2
27.Hexyl isodecyl phthalate
28.Benzyl 2-ethylhexyl phthalate
29.Bis(4-methylcyclohexyl) phthalate isomer 3
30.Bis(2-ethylhexyl) isophthalate
31.Bis(2-(ethoxyethoxy)ethyl) phthalate
32.Di-n-octyl phthalate
33.n-Hexyl decyl phthalate
34.Diphenyl isophthalate
35.Dibenzyl phthalate
36.Diisononyl phthalate
37.Di-n-octyl isophthalate
38.Diisodecyl phthalate
39.Dinonyl phthalate
40.n-Octyl n-decyl phthalate
image
GC_EV1407
ColumnRtx-440, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12923)
SamplePhthalate standards mixture
Benzyl benzoate (cat.# 31847)
Diluent:Methylene chloride
Conc.:50 µg/mL (80 μg/mL for internal standard benzyl benzoate)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 3.5 mm Precision liner w/wool (cat.# 23320.1)
Inj. Temp.:280 °C
Split Vent Flow Rate:3 mL/min
Oven
Oven Temp.:150 °C (hold 0.8 min) to 200 °C at 5 °C/min to 275 °C at 3 °C/min (hold 2 min)
Carrier GasHe, constant linear velocity
Linear Velocity:48 cm/sec, 21.4 psi, 147.5 kPa @ 150 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.559-400-
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Electron Energy:70 eV
Solvent Delay Time:2.5 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesThe constant linear velocity of 48 cm/sec is equivalent to approximately 1.61 mL/min at 150 °C. The MS scan interval is 0.1 sec.
AcknowledgementThe authors would like to thank Shimadzu Corporation for their consultation with the operation of the QP2010 Plus GC-MS instrument.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

selectivity

,

Rtx-440

,

GC-MS

,

Phthalates

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us