share:
delim
delim
Resolution of Indeno[123-cd]pyrene from Dibenz Anthracene Isomers on Rxi-PAH (60 m x 0.25 mm x 0.10 µm)

Resolution of Indeno[123-cd]pyrene from Dibenz Anthracene Isomers on Rxi-PAH (60 m x 0.25 mm x 0.10 µm)

PeakstR (min)Conc.
(µg/mL)
Quant Ion
1.Dibenzo[a,j]anthracene74.462.0278
2.Dibenzo[a,c]anthracene75.181.3278
3.Indeno[1,2,3-cd]pyrene75.318.8276
4.Dibenzo[a,h]anthracene75.398.8278
5.Benzo[b]chrysene76.111.8278
6.Picene76.451.4278
7.Benzo[ghi]perylene77.089.3276
8.Anthanthrene77.870.95276
image
GC_EV1424
ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
SampleSV internal standard mix (cat.# 31206)
Coronene-D12 (CIL) (cat.# DLM-2715)
Benzo[a]pyrene-D12 (CIL) (cat.# DLM-258-0)
Aromatics in toluene (NIST) (cat.# 2260a)
Native PAH stock (Wellington Labs) (cat.# PAH-STK-A)
EU 15+1 PAH standard (cat.# 32470)
Custom PAH SIM standard (cat.# 557484)
Diluent:Toluene
Conc.:0.71 to 10 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:110 °C (hold 1.6 min) to 175 °C at 30 °C/min to 265 °C at 1.6 °C/min to 350 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
15.09102.1, 108.1, 128.1, 136.2 20
26.68115.1, 142.1 20
37.5576.1, 141.1, 154.1, 156.2 20
48.7775.6, 152.1 20
59.5676.1, 153.1, 162.2, 164.2 20
610.26155.1, 170.2 20
711.1282.4, 165.1 20
814.27139.1, 184.1 20
917.26152.1, 160.2, 178.1, 188.2 20
1020.5894.6, 165.1, 190.1, 192.1 20
1126.32101.1, 202.1 20
1234.9792.1, 184.1 20
1337.36108.0, 216.0 20
1441.78196.1, 212.2 20
1545.86113.1, 226.1, 228.1 20
1647.86114.0, 228.1 20
1749.49113.1, 120.1, 226.1, 228.1, 240.1 20
1851.24154.1, 252.1 20
1954.16119.8, 242.2 20
2060.13125.1, 126.1, 252.1 20
2164.77126.1, 252.1 20
2266.15125.1, 126.1, 132.1, 252.1, 264.1 20
2368.07125.0, 132.2, 252.1, 264.1 20
2469.10252.1, 268.1 20
2571.92139.1, 139.5, 278.1, 279.1 20
2674.87138.1, 139.1, 276.1, 278.1 20
2775.81138.1, 139.1, 278.1 20
2876.82138.1, 276.1 20
2977.53132.6, 138.1, 267.1, 276.1 20
3080.20151.0, 302.1 20
3183.40150.0, 151.0, 156.1, 300.1, 302.1, 312.1 20
3284.88151.0, 302.1 20
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:3 min
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesConditions optimized using EZGC software produce good separation of dibenzo[a,c]anthracene and dibenzo[a,h]anthracene from indeno[1,2,3-cd]pyrene, triphenylene from chrysene, as well as the benzofluoranthene isomers.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAH

,

Indenopyrene

,

dibenzanthracene

,

benzofluoranthenes

,

EFSA 15+1

,

16 priority pollutant PAHs

,

EFSA 4

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us