share:
delim
delim
Nonylphenol, Bisphenol A, 4-tert-Octylphenol, Nonylphenol Monoethoxylate, and Nonylphenol Diethoxylate on the Rxi-5Sil MS by ASTM D7065-11 (Constant Flow)
This chromatogram may contain analytes that are synonyms of your search terms.

Nonylphenol, Bisphenol A, 4-tert-Octylphenol, Nonylphenol Monoethoxylate, and Nonylphenol Diethoxylate on the Rxi-5Sil MS by ASTM D7065-11 (Constant Flow)

PeakstR (min)Conc.
(µg/mL)
On-Column (ng)
1.Acenaphthene-D1013.6252.5
2.4-tert-Octylphenol15.4323.2
3.Nonylphenol15.9–17.016016
4.Phenanthrene-D1017.1252.5
5.4-Nonylphenol18.0323.2
6.Nonylphenol monoethoxylate18.5–20.032032
7.4-Nonylphenol monoethoxylate20.5323.2
8.Bisphenol A20.9323.2
9.Nonylphenol diethoxylate21.0–22.364064
10.Chrysene-D1223.4252.5
image
GC_EV1426
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMethod 525.2 internal standard mix (cat.# 31825)
AccuStandard nonylphenol calibration standard (cat.# M-1626)
Diluent:Dichloromethane
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:290 °C
Oven
Oven Temp.:50 °C (hold 2 min) to 320 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.44 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
13.045-5000.8
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9mm ID
Source Temp.:325 °C
Quad Temp.:150 °C
Solvent Delay Time:3.0 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesThe use of constant flow, rather than constant pressure, yields a faster analysis and more constant peak widths throughout the chromatogram.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us