share:
delim
delim
EPA Method 530 UCMR4 Standard at 10x the Method Reporting Limit on Rtx-1701 (SIM)
This chromatogram may contain analytes that are synonyms of your search terms.

EPA Method 530 UCMR4 Standard at 10x the Method Reporting Limit on Rtx-1701 (SIM)

PeakstR (min)Conc.
(µg/mL)
1.o-Toluidine-d9 (SS)8.431.0
2.o-Toluidine8.480.70
3.Quinoline-d7 (SS)10.621.0
4.Quinoline10.642.0
5.Acenaphthene-d10 (IS)13.451.0
6.Butylated hydroxyanisole (BHA)14.253.0
7.Phenanthrene-d10 (IS)17.321.0
image
GC_EV1467
ColumnRtx-1701, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12023)
SampleMethod 530 UCMR4 standard (cat.# 572262)
Method 530 UCMR4 surrogate standard (cat.# 572265)
Method 530 UCMR4 internal standard (cat.# 572266)
Diluent:Dichloromethane
Injection
Inj. Vol.:1 µL pulsed splitless (hold 1.0 min)
Liner:Topaz 4 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:275 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:1.05 min
Purge Flow:60 mL/min
Oven
Oven Temp.:60 °C (hold 1 min) to 290 °C at 10 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
13.50106, 107, 112, 114 25
29.55102, 108, 129, 136 25
312.08162, 164 25
414.02137, 180 25
515.92160, 188 25
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Stainless Steel
Drawout Plate:6 mm ID
Source Temp.:280 °C
Quad Temp.:180 °C
Solvent Delay Time:1.45 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentHP6890 GC & 5973 MSD
NotesThe EPA Method 530 UCMR4 standard analyte concentrations vary to simplify preparing ICAL levels based on the minimum method reporting levels.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us