share:
delim
delim
EPA Method 541 UCMR4 Standard at 10x the Method Reporting Limit on Stabilwax (SIM)
This chromatogram may contain analytes that are synonyms of your search terms.

EPA Method 541 UCMR4 Standard at 10x the Method Reporting Limit on Stabilwax (SIM)

PeakstR (min)Conc.
(µg/mL)
1.2-Propen-1-ol6.940.13
2.1-Butanol-d10 (SS)7.250.25
3.1-Butanol7.380.50
4.2-Methoxyethanol7.970.10
5.Chlorobenzene-d5 (IS)8.660.25
image
GC_EV1469
ColumnStabilwax, 30 m, 0.25 mm ID, 0.50 µm (cat.# 10638)
SampleMethod 541 UCMR4 standard (cat.# 572263)
Method 541 UCMR4 internal standard (cat.# 572268)
Method 541 UCMR4 surrogate standard (cat.# 572267)
Diluent:Dichloromethane
Injection
Inj. Vol.:1 µL pulsed splitless (hold 0.50 min)
Liner:Topaz 4 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:200 °C
Pulse Pressure:10 psi (68.9kPa)
Pulse Time:0.55 min
Purge Flow:100 mL/min
Oven
Oven Temp.:30 °C (hold 0.5 min) to 110 °C at 10 °C/min to 200 °C at 25 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:0.9 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
16.039, 55, 57, 58 25
27.241, 43, 46, 50, 55, 56, 63, 64 25
37.745, 47, 58, 76 25
48.376, 117, 119 25
Transfer Line Temp.:200 °C
Analyzer Type:Quadrupole
Source Type:Stainless Steel
Drawout Plate:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Solvent Delay Time:6.0 min
Tune Type:BFB
Ionization Mode:EI
InstrumentHP6890 GC & 5973 MSD
NotesThe 30 m x 0.25 mm ID x 0.5 µm Stabiliwax column offers equivalent selectivity to the ZB-Wax Plus chromatogram published in EPA Method 541, but with a significantly faster analysis time under the same GC conditions. Chlorobenzene elutes at 13.14 min in the method and at 8.66 minutes on the Stabiliwax column.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us