share:
delim
delim
MA DEP EPH Aliphatics on Rxi-5Sil MS by Split Injection

MA DEP EPH Aliphatics on Rxi-5Sil MS by Split Injection

PeaksConc.
(µg/mL)
1.n-Nonane2.40
2.n-Decane3.14
3.n-Dodecane4.64
4.n-Tetradecane5.97
5.n-Hexadecane6.98
6.n-Octadecane7.80
7.n-Nonadecane8.17
PeaksConc.
(µg/mL)
8.n-Eicosane8.51
9.1-Chlorooctadecane (SS)8.81
10.n-Docosane9.20
11.n-Tetracosane10.05
12.n-Hexacosane11.15
13.n-Octacosane12.56
14.n-Triacontane14.24
15.n-Hexatriacontane17.15
image
GC_EV1488
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMA EPH aliphatic hydrocarbon standard (cat.# 31459)
1-Chlorooctadecane (cat.# 31098)
Diluent:Hexane
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:300 °C
Oven
Oven Temp.:55 °C (hold 0.5 min) to 150 °C at 20 °C/min to 250 °C at 30 °C/min to 280 °C at 5 °C/min to 345 °C at 30 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us