NJ DEP EPH Aromatics on Rxi-5Sil MS by Split Injection
This chromatogram may contain analytes that are synonyms of your search terms.

NJ DEP EPH Aromatics on Rxi-5Sil MS by Split Injection

PeakstR (min)Conc.
(µg/mL)
1.1,2,3-Trimethylbenzene3.33125
2.Naphthalene4.60125
3.2-Methylnaphthalene5.40125
4.2-Fluorobiphenyl (SS)5.82200
5.Acenaphthylene6.34125
6.2-Bromonaphthalene (SS)6.48200
7.Acenaphthene6.52125
8.Fluorene7.01125
9.Phenanthrene7.86125
10.Anthracene7.90125
11.o-Terphenyl (SS)8.19200
12.Fluoranthene8.88125
13.Pyrene9.09125
14.Benz[a]anthracene10.57125
15.Chrysene10.63125
16.Benzo[b]fluoranthene12.55125
17.Benzo[k]fluoranthene12.61125
18.Benzo[a]pyrene13.23125
19.Indeno[1,2,3-cd]pyrene15.57125
20.Dibenz[a,h]anthracene15.64125
21.Benzo[ghi]perylene15.92125
image
GC_EV1489
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleNJ DEP EPH 10/08 rev. 2 aromatics calibration standard (cat.# 30541)
MA fractionation surrogate spike mix (cat.# 31480)
o-Terphenyl (cat.# 31097)
Diluent:Methylene chloride
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:300 °C
Oven
Oven Temp.:55 °C (hold 0.5 min) to 150 °C at 20 °C/min to 250 °C at 30 °C/min to 280 °C at 5 °C/min to 345 °C at 30 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us