share:
delim
delim
TNRCC 1006 Retention Time Marker Mix on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

TNRCC 1006 Retention Time Marker Mix on Rxi-5Sil MS

PeakstR (min)Conc.
(µg/mL)
1.n-Hexane1.31200
2.n-Heptane2.09200
3.n-Octane3.32200
4.n-Decane5.46200
5.n-Dodecane7.11200
6.n-Hexadecane9.74200
7.n-Heneicosane12.33200
8.n-Octacosane15.14200
9.n-Pentatriacontane17.43200
image
GC_EV1492
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleTNRCC 1006 retention time marker mix (cat.# 31814)
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Topaz 4 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:300 °C
Oven
Oven Temp.:35 °C (hold 2 min) to 330 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us