share:
delim
delim
Resprep CarboPrep Plus SPE Cartridge (Lot 1226 Comparison)
This chromatogram may contain analytes that are synonyms of your search terms.

Resprep CarboPrep Plus SPE Cartridge (Lot 1226 Comparison)

Peaks
1.2,4,5,6-Tetrachloro-m-xylene (TCMX)
Peaks
2.Decachlorobiphenyl (DCB)
image
GC_EV1494
ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Diluent:Hexane
Conc.:2 ppb
Injection
Inj. Vol.:4 µL splitless (hold 0.75 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:90 °C (hold 0.5 min) to 330 °C at 25 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
DetectorMicro-ECD @ 340 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
NotesConditioned a Resprep CarboPrep Plus SPE cartridge (cat.# 25845) by adding one cartridge volume of hexane:acetone (90:10) and letting it stand for five minutes before drawing the solvent down to frit level. Then, 1 mL of sample extract was loaded onto the cartridge and eluted with 9 mL of hexane:acetone (90:10).
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us