share:
delim
delim
Fast Dual Column Analysis of Organochlorine Pesticides (EPA Method 8081) Using a GC Accelerator Kit and Rtx-CLPesticides and Rtx-CLPesticides2 Column Set
This chromatogram may contain analytes that are synonyms of your search terms.

Fast Dual Column Analysis of Organochlorine Pesticides (EPA Method 8081) Using a GC Accelerator Kit and Rtx-CLPesticides and Rtx-CLPesticides2 Column Set

PeaksConc.
(ng/mL)
1.Tetrachloro-m-xylene25
2.α-BHC25
3.γ-BHC25
4.β-BHC25
5.δ-BHC25
6.Heptachlor25
7.Aldrin25
8.Heptachlor epoxide25
9.trans-Chlordane25
10.cis-Chlordane25
11.4,4'-DDE50
PeaksConc.
(ng/mL)
12.Endosulfan I50
13.Dieldrin50
14.Endrin50
15.4,4'-DDD50
16.Endosulfan II50
17.4,4'-DDT50
18.Endrin aldehyde50
19.Methoxychlor250
20.Endosulfan sulfate50
21.Endrin ketone50
22.Decachlorobiphenyl50
image
GC_EV1498
ColumnsRtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141-051)
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324-051)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with universal angled “Y” Press-Tight connector (cat.# 20403-261)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Pesticide surrogate mix (cat.# 32453)
Diluent:n-Hexane
Injection
Inj. Vol.:2 µL splitless (hold 0.3 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:120 °C (hold 0.3 min) to 200 °C at 60 °C/min to 230 °C at 25 °C/min to 320 °C at 35 °C/min (hold 1.5 min)
Carrier GasH2, constant flow
Linear Velocity:90 cm/sec
DetectorMicro-ECD @ 330 °C
Make-up Gas Flow Rate:60 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesA GC Accelerator oven insert kit (cat.# 23849) was installed in the instrument to speed up analysis. The analytical columns were custom products that were not on cages.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

accelerator

,

organochlorine

,

CLPesticides

,

pesticides

,

8081

,

CLPesticides2

,

fast

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us