Conditioning column above method temperature removes contaminants

Conditioning column above method temperature removes contaminants

image
GC_EX00957
ColumnRxi-5ms, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13424)
Sampleno sample injected; instrument ran the method conditions with no injection.
Injectionsplitless (hold 0.5 min)
Liner:4mm Split w/Wool, IP deactivated (cat.# 20782)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:50 °C (hold 5 min) to 320 °C at 30 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:5.0 mL/min
DetectorFID @ 330 °C
NotesSystem was conditioned for 10min. at the following temperatures: inlet 300°C, oven 310°C, detector 330°C.