Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Peppermint Oil on MXT-WAX
Peaks | |
---|---|
1. | α-Pinene![]() |
2. | β-Pinene![]() |
3. | Sabinene![]() |
4. | Myrcene![]() |
5. | α-Terpinene![]() |
6. | L-Limonene![]() |
7. | 1,8-Cineole![]() |
8. | cis-Ocimene![]() |
9. | γ-Terpinene![]() |
10. | p-Cymene![]() |
11. | Terpinolene![]() |
12. | 3-Octanol![]() |
13. | 1-Octen-3-ol![]() |
14. | l-Menthone![]() |

GC_FF00141
Column | MXT-WAX, 30 m, 0.28 mm ID, 0.50 µm (cat.# 70639) |
Sample | peppermint oil |
Conc.: | neat |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 50:1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 75 °C (hold 4 min) to 240 °C at 4 °C/min |
Carrier Gas | H2, constant pressure |
Linear Velocity: | 40 cm/sec @ 75 °C |
Detector | FID @ 250 °C |
Notes | FID sensitivity: 16 x 10-11 AFS |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.