share:
delim
delim
FAMEs (Saw Palmetto) on Stabilwax

FAMEs (Saw Palmetto) on Stabilwax

Peaks
1.Caproic acid
2.Caprylic acid
3.Pelargonic acid
4.Capric acid
5.Lauric acid
6.Myristic acid
7.Palmitic acid
8.Margaric acid
9.Stearic acid
10.Oleic acid
11.Linoleic acid
12.Linolenic acid
image
GC_FF00367
ColumnStabilwax, 50 m, 0.25 mm ID, 0.25 µm (cat.# 10626-105)
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:230 °C
Split Vent Flow Rate:37.5 mL/min
Oven
Oven Temp.:110 °C (hold 1 min) to 240 °C at 8 °C/min (hold 25 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 250 °C
NotesSeptum purge: 3 mL/min
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us