share:
delim
delim
Food Packaging Volatiles by Purge & Trap GC/MS on Rtx-5MS, Overwrap at the Seam
This chromatogram may contain analytes that are synonyms of your search terms.

Food Packaging Volatiles by Purge & Trap GC/MS on Rtx-5MS, Overwrap at the Seam

Peaks
1.Tetrahydrofuran
2.1-Butanol
3.Toluene
4.Hexanal
5.Ethylbenzene
6.4-Heptanone
7.Butyl ether
Peaks
8.Styrene
9.1-Methylethylbenzene
10.Propylbenzene
11.Benzaldehyde
12.Benzeneacetaldehyde
13.Acetophenone
14.Benzoic acid, methyl ester
15.Decanal
image
GC_FF00459
ColumnRtx-5MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 12653)
Injectionpurge and trap split (split ratio 20:1)
Liner:1 mm ID
Inj. Temp.:250 °C
Purge and Trap
Instrument:Tekmar LSC-3100 purge & trap
Trap Type:Vocarb 3000, type K
Purge: 10 min @ 60 °C, flow 40 mL/min
Dry Purge: 3 min, flow 40 mL/min
Desorb Preheat Temp.:220 °C
Desorb: 2 min @ 245 °C, flow 40 mL/min
Bake: 6 min @ 230 °C
Interface Connection:injection port
Transfer Line Tubing:Siltek/Sulfinert tubing
Oven
Oven Temp.:50 °C to 92 °C at 3 °C/min to 220 °C at 20 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
Linear Velocity:36 cm/sec
DetectorMS
Mode:Scan
Source Temp.:280 °C
Electron Energy:EI @ 70 eV
Scan Range:35-260 amu
InstrumentHP6890 GC & 5973 MSD
Notescatalog number 22501 Siltek/Sulfinert tubing obsolete, recommend catalog number 29234
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

food contaminants

,

purge and trap

,

P&T

,

purge

,

trap

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us