Alcoholic Beverages (Acids and Esters): Caproic Acid and Ethyl Laurate on Stabilwax-DA
This chromatogram may contain analytes that are synonyms of your search terms.

Alcoholic Beverages (Acids and Esters): Caproic Acid and Ethyl Laurate on Stabilwax-DA

Peaks
1.Caproic acid
2.Ethyl laurate
image
GC_FF00525
ColumnStabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752)
Sample
Diluent:ethyl acetate
Conc.:100 ppm
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:4 mm splitless liner with wool (cat.# 20814-202.1)
Inj. Temp.:240 °C
Oven
Oven Temp.:80 °C to 230 °C at 5 °C/min
Carrier GasH2, constant pressure (28 psi, 193.1 kPa)
Temp.:240 °C
DetectorFID
Make-up Gas Type:N2