Flavor Compounds in Malt Whiskey on Stabilwax-DA

Flavor Compounds in Malt Whiskey on Stabilwax-DA

Peaks
1.Ethyl octanoate
2.Acetic acid
3.Siloxane
4.1-hydroxy-2,3-butadione
5.1-Hydroxy-2-propanone acetate
6.Furfural
7.Formic acid
8.Propionic acid
9.Isobutyric acid
10.Dimethyl Sulfoxide
11.5-Methyl furfural
12.methyl decanoate + unknown
Peaks
13.Butyric acid
14.Siloxane
15.Ethyl decanoate
16.furfuryl alcohol + isoamyl octanoate
17.isovaleric acid + 2-methylbutyric acid
18.Diethyl succinate
19.3-Methyl-2(5H)-furanone (possible)
20.Valeric acid
21.Ethyl undecanoate
22.Isobutyl decanoate
23.2(5H)-Furanone
24.Unknown
25.Diethyl pentanedioate
image
GC_FF00555
ColumnStabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752)
Injection
Inj. Vol.:10 µL splitless
Oven
Oven Temp.:60 °C (hold 2 min) to 100 °C at 20 °C/min to 240 °C at 5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:45 cm/sec
DetectorMS
Mode:Scan
Transfer Line Temp.:240 °C
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Ionization Mode:EI
Scan Range:30-400 amu
NotesInj: 10µL/min
Std. conc.: neat
Gerstel CIS Injector: 35°C (hold 2 min.), to 300°C @ 10°C/sec. (hold 5 min.)
Helium vent flow: 600mL/min with 1.8 min. vent end time
AcknowledgementChromatogram courtesy of Kevin MacNamara, Ph.D., Irish Distilleries, Ltd.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

isoamyl octanoate

,

2-methylbutyric acid

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us