Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Flavor Compounds in Grappa on Stabilwax-DA

GC_FF00560
Column | Stabilwax-DA, 30 m, 0.18 mm ID, 0.18 µm (cat.# 550752) |
Injection | |
Inj. Vol.: | 10 µL splitless |
Oven | |
Oven Temp.: | 60 °C (hold 2 min) to 100 °C at 20 °C/min to 240 °C at 5 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 45 cm/sec |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 240 °C |
Source Temp.: | 230 °C |
Quad Temp.: | 150 °C |
Electron Energy: | 70 eV |
Ionization Mode: | EI |
Scan Range: | 30-400 amu |
Notes | Inj: 10 µL/min Std. conc.: neat Gerstel CIS Injector: 35°C (hold 2 min.), to 300°C @ 10°C/sec. (hold 5 min.) Helium vent flow: 600mL/min with 1.8 min. vent end time |
Acknowledgement | Chromatogram courtesy of Kevin MacNamara, Ph.D., Irish Distilleries, Ltd. |
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.