share:
delim
delim
FAMEs (NLEA Mix) on Rt-2560

FAMEs (NLEA Mix) on Rt-2560

Peaks
1.C4:0 Methyl butyrate
2.C6:0 Methyl hexanoate
3.C8:0 Methyl octanoate
4.C10:0 Methyl decanoate
5.C11:0 Methyl undecanoate
6.C12:0 Methyl laurate
7.C13:0 Methyl tridecanoate
8.C14:0 Methyl myristate
9.C14:1 Methyl myristoleate (cis-9)
10.C15:0 Methyl pentadecanoate
11.C16:0 Methyl palmitate
12.C16:1 Methyl palmitoleate (cis-9)
13.C17:0 Methyl heptadecanoate
14.C18:0 Methyl stearate
Peaks
15.C18:1 Methyl elaidate (trans-9)
16.C18:1 Methyl oleate (cis-9)
17.C18:2 Methyl linolelaidate (trans-9,12)
18.C18:2 Methyl linoleate (cis-9,12)
19.C20:0 Methyl arachidate
20.C20:1 Methyl eicosenoate (cis-11)
21.C18:3 Methyl linolenate (cis-9,12,15)
22.C22:0 Methyl behenate
23.C22:1 Methyl erucate (cis-13)
24.C23:0 Methyl tricosanoate
25.C24:0 Methyl lignocerate
26.C20:5 Methyl eicosapentaenoate (cis-5,8,11,14,17)
27.C24:1 Methyl nervonate (cis-15)
28.C22:6 Methyl docosahexaenoate (cis-4,7,10,13,16,19)
image
GC_FF00651
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13199)
SampleNLEA FAME mix (cat.# 35078)
Diluent:Methylene chloride
Conc.:30 mg/mL total FAMEs
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Splitless (4 mm ID) (cat.# 20814)
Inj. Temp.:225 °C
Oven
Oven Temp.:100 °C (hold 4 min) to 240 °C at 3 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:1.2 mL/min
DetectorFID @ 250 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

NLEA

,

Rt-2560

,

FAMEs

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us