share:
delim
delim
Commercial Peppermint Oil on Rt-γDEXsa

Commercial Peppermint Oil on Rt-γDEXsa

Peaks
1.Menthone
2.Menthol
3.Menthone
4.Menthol
5.Menthyl acetate
image
GC_FF00802
ColumnRt-yDEXsa, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13113)
SampleNeat
Injection
Inj. Vol.:1 µL split (split ratio 150:1)
Inj. Temp.:230 °C
Oven
Oven Temp.:40 °C to 120 °C at 5 °C/min to 135 °C at 3 °C/min to 200 °C at 5 °C/min
Carrier GasHe, constant pressure
Linear Velocity:35 cm/sec @ 100 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:200 °C
Ionization Mode:EI
Scan Range:40-300 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Peppermint

,

Rt-γDEXsm™

,

Chiral

,

Oil

,

chiral gc separation

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us