Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Patchouli Oil Fast GC/MS on Rtx-5
Peaks | |
---|---|
1. | β-Patchoulene![]() |
2. | β-Elemene |
3. | Caryophyllene![]() |
4. | α-Guaiene![]() |
5. | Seychellene![]() |
6. | α-Patchoulene![]() |
7. | Guaiene![]() |
8. | δ-Guaiene![]() |
9. | Selinene |
10. | Patchouli alcohol![]() |

GC_FF00851
Column | Rtx-5, 10 m, 0.10 mm ID, 0.10 µm (cat.# 41201) |
Sample | Patchouli Oil |
Diluent: | Acetone |
Conc.: | 5% |
Injection | |
Inj. Vol.: | 0.2 µL split (split ratio 800:1) |
Inj. Temp.: | 280 °C |
Oven | |
Oven Temp.: | 40 °C to 300 °C at 30 °C/min |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 60 cm/sec, 61.13 psi, 421.5 kPa @ 40 °C |
Detector | MS |
Mode: | Scan |
Source Temp.: | 280 °C |
Ionization Mode: | EI |
Scan Range: | 40-500 amu |
Instrument | Shimadzu 2010 GC & QP2010+ MS |
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.