Underivatized Cholesterol on Rxi-5ms (Temp. program)
This chromatogram may contain analytes that are synonyms of your search terms.

Underivatized Cholesterol on Rxi-5ms (Temp. program)

Peaks
1.5-α-Cholestane (IS)
Peaks
2.Cholesterol
image
GC_FF00881
ColumnRxi -5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420)
Sample1,000 μg/mL cholesterol in DMF
1,000 μg/mL 5-α-cholestane in hexane
25 ng cholesterol, 150 ng 5-α-cholestane on column
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Splitless taper (4 mm) w/wool (cat.# 22405)
Inj. Temp.:250 °C
Oven
Oven Temp.:200 °C (hold 1 min) to 330 °C at 20 °C/min (hold 7.5 min)
Carrier GasHe, constant pressure (9.7 psi, 66.9 kPa)
Linear Velocity:24 cm/sec @ 200 °C
DetectorFID @ 340 °C
Download PDF