share:
delim
delim
Melamine Spike in Cat Food on Rtx-5MS

Melamine Spike in Cat Food on Rtx-5MS

Peaks
1.Cyanuric acid
2.Ammelide
3.Ammeline
4.Melamine
5.Benzoguanamine
image
GC_FF00977
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
SampleMelamine, cyanuric acid, ammelide, ammeline, benzoguanamine in dry cat food
Conc.:10 µg/mL prederivatized
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5 mm splitless taper (cat.# 22286)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:290 °C
Ionization Mode:EI

RELATED CHROMATOGRAM LIBRARY SEARCHES

melamine

,

melamine and analogs

,

cyanuric

,

acid

,

pet food

,

cat food

,

derivatized melamine

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us