share:
delim
delim
Melamine Spike on Rtx-5MS by GC/MS scan

Melamine Spike on Rtx-5MS by GC/MS scan

Peaks
1.Cyanuric acid
2.Ammelide
3.Ammeline
4.Melamine
5.Benzoguanamine
image
GC_FF00979
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Samplemelamine, cyanuric acid, ammelide, ammeline, benzoguanamine
Conc.:10 µg/mL prederivatized
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C (hold 1 min) to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:290 °C
Ionization Mode:EI
Scan Range:50-450 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

adulterants

,

melamine

,

cyanuric acid

,

pet food

,

cat food

,

dog food

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us