share:
delim
delim
Melamine Spike on Rtx-5MS by GC/MS scan
This chromatogram may contain analytes that are synonyms of your search terms.

Melamine Spike on Rtx-5MS by GC/MS scan

Peaks
1.Cyanuric acid
2.Ammelide
3.Ammeline
4.Melamine
5.Benzoguanamine
image
GC_FF00979
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Samplemelamine, cyanuric acid, ammelide, ammeline, benzoguanamine
Conc.:10 µg/mL prederivatized
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C (hold 1 min) to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:290 °C
Ionization Mode:EI
Scan Range:50-450 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

adulterants

,

melamine

,

cyanuric acid

,

pet food

,

cat food

,

dog food

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.