share:
delim
delim
Chlorinated Pesticide Residues in Olive Oil Rtx-CLPesticides2

Chlorinated Pesticide Residues in Olive Oil Rtx-CLPesticides2

PeaksQuant. ionQual. ion 1Qual. ion 2
1.α-BHC*219181109
2.γ-BHC219181109
3.β-BHC219181109
4.δ-BHC219181109
5.Heptachlor272237100
6.Aldrin263293220
7.Heptachlor epoxide26323781
8.δ-Chlordane27223765
9.α-Chlordane27223765
10.Endosulfan I195207241
11.4,4'-DDE246318176
12.Dieldrin79263277
13.Endrin26328181
14.4,4'-DDD235165199
15.Endosulfan II195207-
16.4,4'-DDT235165199
17.Endrin aldehyde67250345
18.Endosulfan sulfate272229239
19.Methoxychlor227274-
20.Endrin ketone67317281
* α-BHC not detected due to low recovery during sample preparation.
image
GC_FF01043
ColumnRtx-CLPesticides2, 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
SampleOlive oil spiked with Organochlorine Pesticide Mix AB # 3 (cat.# 32415)
Conc.:10 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:Single Taper w/Wool (cat.# 22286-200.1)
Inj. Temp.:225 °C
Oven
Oven Temp.:140 °C (hold 0.5 min) to 268 °C at 20 °C/min to 290 °C at 3 °C/min to 330 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:320 °C
Ionization Mode:EI
NotesExtraction and dSPE Cleanup for Pesticide Residues in Olive Oil

Test sample: A 1.5 mL sample of commercially obtained virgin olive oil was spiked with a standard organochlorine pesticide mix. The spiked sample was processed as follows:
1. Dilute with 1.5 mL hexane.
2. Add 6 mL of acetonitrile (ACN).
3. Mix for 30 minutes on a shaker.
4. Allow layers to separate (approximately 20 minutes), then collect the top (ACN) layer.
5. Repeat the liquid-liquid extraction (steps 2–4) and combine both ACN extract layers.
6. Place 1 mL of the combined ACN extract in a 1.5 mL tube containing 150 mg magnesium sulfate and 50 mg PSA.
7. Shake the tube for 2 minutes.
8. Centrifuge at 3,000 U/min for approximately 5 minutes.
9. Remove the top layer and inject directly into the gas chromatograph system.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

olive oil

,

OCP

,

pesticides

,

CLP

,

chlorinated pesticides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us