share:
delim
delim
Analysis of Melamine and Related Compounds on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Analysis of Melamine and Related Compounds on Rxi-5Sil MS

PeakstR (min)Quant. IonQual. IonQual.IonQual. Ion
1.Cyanuric acid10.238345330346347
2.Ammelide11.090344329345330
3.Ammeline11.770328343329344
4.Melamine12.318327342328343
5.Benzoguanamine14.553316331332330
image
GC_FF01094
ColumnRxi-5Sil MS w/5 m Integra-Guard column, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623-124)
SampleMelamine and related analogs stock standard (cat.# 33253)
Benzoguanamine (cat.# 33251)
Conc.:0.01 µg/mL as tri-TMS derivatives
Injection
Inj. Vol.:1.0 µL splitless (hold 1 min)
Liner:3.5 mm Gooseneck Splitless w/Wool (cat.# 22286-200.1)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:290 °C
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
Notesall method ions in table, only quantification ions were plotted
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

infant formula

,

melamine

,

pet food

,

cyanuric

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us