share:
delim
delim
Polycyclic Aromatic Hydrocarbons Infant Formula with and without QuEChERS dSPE Clean-up on Rxi-5Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

Polycyclic Aromatic Hydrocarbons Infant Formula with and without QuEChERS dSPE Clean-up on Rxi-5Sil MS

Peaksm/z
1.decafluorobiphenyl (SS)265
2.Naphthalene128
3.2-Methylnaphthalene142
4.1-Methylnaphthalene142
5.Acenaphthylene152
6.Acenaphthene152
7.Fluorene166
8.phenanthrene-d10 (IS)188
9.Phenanthrene178
10.Anthracene178
Peaksm/z
11.Fluoranthene202
12.Pyrene202
13.Benzo[a]anthracene228
14.Chrysene228
15.Benzo[b]fluoranthene252
16.Benzo[k]fluoranthene252
17.Benzo[a]pyrene252
18.Indeno[1,2,3-cd]pyrene276
19.Dibenz[a,h]anthracene278
20.Benzo[ghi]perylene276
image
GC_FF01123
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleLiquid infant formula spiked with: (see details in notes section)
decafluorobiphenyl (cat.# 31842)
EPA Method 8310 PAH Mixture (cat.# 31841)
phenanthrene-d10 (cat.# 31045)
CHROMATOGRAM TRACES
A. extract with clean-up using Q-sep Q210 QuEChERS dSPE clean-up tube (cat.# 26215)
B. extract without dSPE clean-up
Injection
Inj. Vol.:1 µL splitless (hold 0.15 min)
Liner:3.5mm Gooseneck Splitless w/Wool (cat.# 22286-200.1)
Inj. Temp.:300 °C
Oven
Oven Temp.:50 °C (hold 0.5 min) to 290 °C at 25 °C/min to 320 °C at 5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesSample Details:
Liquid infant formula spiked at 1μg/mL with decafluorobiphenyl (cat.# 31842) and EPA Method 8310 PAH Mixture (cat.# 31841) and at 0.5μg/mL with internal standard phenanthrene-d10 (cat.# 31045), then extracted with acetonitrile and Q-sep Q110 QuEChERS extraction tube (cat.# 26213)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Formula

,

Polycyclic

,

Aromatic

,

Hydrocarbons

,

Infant

,

QuEChERS

,

dSPE

,

Clean-up

,

Rxi®-5Sil

,

Rxi-5SilMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us