Pure Chromatography
www.restek.com
www.restek.com



This chromatogram may contain analytes that are synonyms of your search terms.
Dandelion Root QuEChERS Extract Cleaned with Cartridge SPE (GCxGC-TOFMS Contour Plot)

GC_FF1189
Column | Rxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
Rtx-200 1.5 m, 0.18 mm ID, 0.20 µm (cat.# 45001) | |
Sample | |
Diluent: | Toluene |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Liner: | Gooseneck Splitless (4mm) w/Wool (cat.# 22405) |
Inj. Temp.: | 250 °C |
Purge Flow: | 40 mL/min |
Oven | |
Oven Temp.: | Rxi-5Sil MS: 80 °C (hold 1 min) to 310 °C at 4 °C/min (hold 1.5 min) Rtx-200: 90 °C (hold 1 min) to 320 °C at 4 °C/min (hold 1.5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.8 mL/min |
Modulation | |
Modulator Temp. Offset: | 25 °C |
Second Dimension Separation Time: | 4 sec |
Hot Pulse Time: | 1.2 sec |
Cool Time between Stages: | 0.8 sec |
Detector | TOFMS |
Transfer Line Temp.: | 290 °C |
Analyzer Type: | TOF |
Source Temp.: | 225 °C |
Electron Energy: | 70 eV |
Mass Defect: | -20 mu/100 u |
Solvent Delay Time: | 4 min |
Ionization Mode: | EI |
Acquisition Range: | 45 to 550 amu |
Spectral Acquisition Rate: | 100 spectra/sec |
Instrument | LECO Pegasus 4D GCxGC-TOFMS |
Notes | See application note GNAN1338 for extraction and cleanup details. A 1.5 m length of the Rtx-200 column was trimmed from a 10 m column. Columns were connected with a Universal Press-Tight Connector (cat.# 20429). |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.