GCxGC-TOFMS Mass Spectrum of Fluoranthene from QuEChERS Extract of NIST SRM 2974a Freeze-Dried Mussel Tissue
This chromatogram may contain analytes that are synonyms of your search terms.

GCxGC-TOFMS Mass Spectrum of Fluoranthene from QuEChERS Extract of NIST SRM 2974a Freeze-Dried Mussel Tissue

image
GC_FF1198
ColumnRxi-17Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 14123)
Rxi-1ms 1.2 m, 0.15 mm ID, 0.15 µm (cat.# custom)
Sample
Diluent:Acetonitrile
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Gooseneck Splitless (4mm) w/Semivolatiles Wool (cat.# 20798-231.1)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:Rxi-17Sil MS: 90 °C (hold 1 min) to 320 °C at 3.75 °C/min (hold 2.67 min)
Rxi-1ms: 95 °C (hold 1 min) to 325 °C at 3.75 °C/min (hold 2.67 min)
Carrier GasHe, corrected constant flow (2.2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:0.9 sec
Cool Time between Stages:0.6 sec
DetectorTOFMS
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:5 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us