share:
delim
delim
Extracted Ion Chromatograms and Spectra for Incurred Pesticides in Strawberries on Rxi-5Sil MS

Extracted Ion Chromatograms and Spectra for Incurred Pesticides in Strawberries on Rxi-5Sil MS

image
GC_FF1201
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleStrawberries processed with QuEChERS sample preparation
QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267)
anthracene for quality control (cat.# 33264)
Diluent:acetonitrile
Conc.: captan incurred in store bought strawberries at 1,400 ppb and thiabendazole incurred in Amish strawberries at about 10 ppb
Injection
Inj. Vol.:1 µL splitless (hold 1.5 min)
Liner:Gooseneck Splitless (5mm) (cat.# 22973-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1.5 min) to 340 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorMS
Mode:
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSample Preparation:
Sample: 10 g of homogenized strawberries, either Amish or store bought, and QuEChERS Internal Standard Mix for GC/MS Analysis (cat.# 33267).

Extraction: 10 mL acetonitrile added, 1 minute shake, then addition of Q-sep Q110 (cat.# 26235), 1 minute shake, centrifuge with Q-sep 3000 Centrifuge (cat.# 26230).

dSPE: Cleanup procedure according to EN 15662 method, add control standard anthracene (cat.# 33264) to 1 mL extract, add this to Q-sep Q210 dSPE tubes (cat.# 26215), shake, centrifuge.

Analysis: Extracted ion chromatograms of m/z 79 and 149 for captan and m/z 174 and 201 for thiabendazole.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

strawberries

,

thiabendazole

,

captan

,

pesticides

,

QuEChERS

,

Q-sep

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us