Marijuana Pesticides by GCxGC on Rxi-5Sil MS and Rtx-200
This chromatogram may contain analytes that are synonyms of your search terms.

Marijuana Pesticides by GCxGC on Rxi-5Sil MS and Rtx-200

PeaksRT 1 (sec)RT 2 (sec)
1.Methamidophos4802.51
2.Dichlorvos5072.11
3.1,2,3,5-Tetrachlorobenzene6211.12
4.Nicotine6541.39
5.Mevinphos7682.59
6.Acephate7710.17
7.Pentachlorobenzene8851.16
8.o-Phenylphenol8881.21
9.Tebuthiuron8942.06
10.Tetrachloronitrobenzene (Tecnazene)9841.59
11.Omethoate9900.25
12.2,3,5,6-Tetrachloroaniline10291.28
13.α-BHC11281.40
14.Hexachlorobenzene11341.16
15.Pentachloroanisole11521.26
16.Dimethoate11642.46
17.β-BHC11911.64
18.Pentachloronitrobenzene11971.52
19.Pentachlorobenzonitrile12061.69
20.γ-BHC12121.43
21.Chlorothalonil12512.56
22.Anthracene12571.42
23.Diazinon12601.25
24.δ-BHC12811.49
25.Pentachloroaniline13141.30
26.Vinclozolin13712.08
27.Carbaryl13832.06
28.Metalaxyl13981.87
29.Pentachlorothioanisole14311.20
30.Pirimiphos methyl14341.33
31.Methiocarb14371.59
32.Dichlofluanid14462.00
33.Malathion14611.79
34.Chlorpyrifos14731.38
35.Fenthion14791.43
36.DCPA methyl ester (Chlorthal-dimethyl)14821.71
37.Parathion14882.10
38.Cyprodinil15481.20
39.Heptachlor epoxide15571.38
40.Thiabendazole15721.47
41.Captan15781.83
42.Folpet15901.74
43.Procymidone15902.24
44.Endosulfan I16381.41
45.Imazalil16741.88
46.4,4'-DDE16951.17
47.Dieldrin16981.40
48.Myclobutanil17072.22
49.Endrin17401.44
50.Endosulfan II17671.55
51.4,4'-DDD17851.31
52.Oxadixyl17852.94
53.2,4'-DDT17911.16
54.Carfentrazone ethyl18481.87
55.Endosulfan sulfate18512.08
56.Fenhexamid18631.44
57.4,4'-DDT18691.23
58.Propargite19111.31
59.Piperonyl butoxide19261.06
60.Iprodione19621.74
61.Bifenthrin19861.17
62.Dicofol19951.20
63.Fenpropathrin20041.51
64.Phosalone20461.86
65.Azinphos-methyl20491.80
66.cis-Permethrin21901.14
67.Coumaphos21962.33
68.trans-Permethrin22051.18
69.Cypermethrin 122921.49
70.Cypermethrin 223041.54
71.Cypermethrin 323101.53
72.Cypermethrin 423161.58
73.Pyraclostrobin24001.42
74.Fluvalinate 124091.72
75.Fluvalinate 224181.70
76.Difenoconazole 124391.55
77.Difenoconazole 224451.55
78.Deltamethrin24781.49
79.Azoxystrobin24961.70
image
GC_FF1205
ColumnRxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Rtx-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15020)
Sample
Diluent:Toluene
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.1)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:Rxi-5Sil MS: 80 °C (hold 1 min) to 310 °C at 5 °C/min
Rtx-200: 85 °C (hold 1 min) to 315 °C at 5 °C/min
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:0.9 sec
Cool Time between Stages:0.6 sec
DetectorTOFMS
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:5 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesRtx-200 (cat.# 15020) is a 15 m column. A 1.3 m section was used as the second dimension column.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

QuEChERS

,

Pesticides

,

Medical marijuana

,

Marijuana

,

GCxGC

,

Cannabis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us