share:
delim
delim
NIST SRM 2260a PAH Mix on Rxi-PAH

NIST SRM 2260a PAH Mix on Rxi-PAH

PeaksQuant Ion (m/z)SIM Group #
1.Naphthalene1281
2.Biphenyl1541
3.Acenaphthylene1521
4.Acenaphthene1531
5.Fluorene1651
6.Dibenzothiophene1842
7.Phenanthrene1782
8.Anthracene1782
9.4H-Cyclopenta[def]phenanthrene1902
10.Fluoranthene2022
11.Pyrene2022
12.Benzo[ghi]fluoranthene2263
13.Benzo[c]phenanthrene2283
14.Benz[a]anthracene2283
15.Cyclopenta[cd]pyrene2263
16.Triphenylene2283
17.Chrysene2283
18.Benzo[b]fluoranthene2524
19.Benzo[k]fluoranthene2524
20.Benzo[j]fluoranthene2524
21.Benzo[a]fluoranthene2524
22.Benzo[e]pyrene2524
23.Benzo[a]pyrene2524
24.Perylene2524
25.Dibenz[a,j]anthracene2785
26.Dibenz[a,c]anthracene2785
27.Indeno[1,2,3-cd]pyrene2765
28.Dibenz[a,h]anthracene2785
29.Benzo[b]chrysene2785
30.Picene2785
31.Benzo[ghi]perylene2765
32.Anthanthrene2765
33.Dibenzo[b,k]fluoranthene3026
34.Dibenzo[a,e]pyrene3026
35.Coronene3006
36.Dibenzo[a,h]pyrene3026
image
GC_FF1223
ColumnRxi-PAH, 40 m, 0.18 mm ID, 0.07 µm (cat.# 49316)
SampleNIST SRM 2260a PAH mix
Diluent:Toluene
Conc.:0.2 - 2 µg/mL (SRM 2260a PAH mix was diluted 5x in toluene)
Injection
Inj. Vol.:0.5 µL pulsed splitless (hold 0.58 min)
Liner:Premium 2 mm single taper w/wool (cat.# 23316.1)
Inj. Temp.:275 °C
Pulse Pressure:80 psi (551.6kPa)
Pulse Time:0.6 min
Purge Flow:40 mL/min
Oven
Oven Temp.:110 °C (hold 1 min) to 210 °C at 37 °C/min to 260 °C at 3 °C/min to 350 °C at 11 °C/min (hold 4.5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
13.00128, 152, 153, 154, 165 40
25.50178, 184, 190, 202 50
313.00226, 228 100
420.00252 200
526.00276, 278 100
629.00300, 302 150
Transfer Line Temp.:350 °C
Analyzer Type:Quadrupole
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:3.00 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAH

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us