Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Two Brands of Gin on Stabilwax-MS (Overlay)
* Not found in gin represented by red trace.

GC_FF1237
Column | Stabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673) | ||||||||
Sample | Two different brands of gin | ||||||||
Conc.: | Neat | ||||||||
Injection | |||||||||
Inj. Vol.: | 1 µL split (split ratio 20:1) | ||||||||
Liner: | Premium 3.5 mm Precision liner w/wool (cat.# 23320.1) | ||||||||
Inj. Temp.: | 250 °C | ||||||||
Oven | |||||||||
Oven Temp.: | 35 °C (hold 5 min) to 250 °C at 7 °C/min (hold 5 min) | ||||||||
Carrier Gas | He, constant linear velocity | ||||||||
Linear Velocity: | 36 cm/sec | ||||||||
Detector | MS | ||||||||
Mode: | Scan | ||||||||
Scan Program: |
| ||||||||
Transfer Line Temp.: | 260 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 250 °C | ||||||||
Solvent Delay Time: | 0.5 min | ||||||||
Ionization Mode: | EI | ||||||||
Instrument | Shimadzu 2010 GC & QP2010+ MS |