Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Cannabis Potency Compounds on Rxi-5Sil MS
Peaks | |
---|---|
1. | Cannabidiol![]() |
2. | Cannabichromene![]() |
3. | delta-8-Tetrahydrocannabinol![]() |
Peaks | |
---|---|
4. | delta-9-Tetrahydrocannabinol![]() |
5. | Cannabigerol![]() |
6. | Cannabinol![]() |

GC_FF1253
Column | Rxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620) |
Sample | Cannabinoids standard (cat.# 34014) |
Cannabichromene (cat.# 34092) | |
delta-8-Tetrahydrocannabinol (THC) (cat.# 34090) | |
Cannabigerol (cat.# 34091) | |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Premium 4 mm Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 150 °C (hold 0.1 min) to 330 °C at 25 °C/min (hold 0.7 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.6 mL/min |
Detector | FID @ 350 °C |
Constant Column + Constant Make-up: | 50 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.