share:
delim
delim
Electronic Cigarette Liquid on Rtx-VMS (GC-MS)
This chromatogram may contain analytes that are synonyms of your search terms.

Electronic Cigarette Liquid on Rtx-VMS (GC-MS)

PeakstR (min)Match QualityEC LiquidBlankRegion
1.Nitrogen/oxygen/carbon dioxide1.051100xxRed
2.Water1.441100xxRed
3.Methanol1.709100xxRed
4.Unidentified1.934xxRed
5.cis-1,2-Dimethylcyclopropane2.11794xxRed
6.Ethanol2.239100xRed
7.1,1-Dichloroethene2.28294xxRed
8.Methylene chloride2.757100xxRed
9.1,2-Dichloroethene2.89194xxRed
10.Ethyl acetate4.03791xRed
11.Unidentified6.000xRed
12.Unidentified6.085xRed
13.Toluene6.207100xRed
14.Propylene glycol7.853100xOrange
15.2,3-Dimethylpyrazine9.24391xOrange
16.Unidentified9.615xOrange
17.Unidentified9.713xOrange
18.Unidentified9.889xOrange
19.Unidentified10.017xOrange
20.Unidentified10.060xOrange
21.Trimethylpyrazine10.38394xOrange
22.Unidentified10.828xOrange
23.Unidentified10.907xOrange
24.Unidentified11.047xOrange
25.Unidentified11.114xOrange
26.Acetylpyrazine11.39495xOrange
27.N-(1-Methylethyl)benzenamine11.86480xOrange
28.Dipropylene glycol12.07191xOrange
29.Glycerin12.473100xOrange
30.Dipropylene glycol methyl ether13.04080xGreen
31.Unidentified13.107xGreen
32.Unidentified13.168xGreen
33.Unidentified13.229xGreen
34.1-(3-Pyridinyl)ethanone13.32194xGreen
35.Unidentified13.412xGreen
36.Unidentified13.463xGreen
37.Unidentified14.479xGreen
38.Unidentified14.534xGreen
39.Unidentified14.643xGreen
40.Unidentified14.863xGreen
41.Unidentified15.003xGreen
42.Nicotine15.800100xGreen
43.Unidentified16.161xBlue
44.Unidentified16.222xBlue
45.α-Damascone16.28995xBlue
46.Unidentified16.374xBlue
47.Unidentified16.417xBlue
48.Unidentified16.478xBlue
49.Unidentified16.643xBlue
50.Unidentified16.984xBlue
51.Unidentified17.033xBlue
52.Myosmine17.15595xBlue
53.Unidentified17.276xBlue
54.Unidentified17.380xBlue
55.Unidentified17.441xBlue
56.Nicotine 1-N-oxide17.53393xBlue
57.Anabasine17.69798xBlue
58.Nicotyrine17.75291xBlue
59.Unidentified18.105xBlue
60.2,3-Dipyridyl18.55097xBlue
61.Unidentified19.788xBlue
62.Unidentified21.025xBlue
63.Unidentified21.092xBlue
64.Cotinine21.63591xBlue
image
GC_FF1260
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample
Diluent:Methylene chloride
Conc.: Electronic cigarette liquid diluted 2:1
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 1 min) to 250 °C at 11 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
Linear Velocity:51.15 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1015-5505.2
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us