share:
delim
delim
Nutritional Labeling FAMEs on Rt-2560 by AOAC Method 996.06

Nutritional Labeling FAMEs on Rt-2560 by AOAC Method 996.06

PeakstR (min)Conc.
(µg/mL)
Structural Nomenclature
1.Methyl butyrate13.1640C4:0
2.Methyl caproate15.5340C6:0
3.Methyl octanoate19.4340C8:0
4.Methyl decanoate24.5040C10:0
5.Methyl undecanoate27.2020C11:0
6.Methyl dodecanoate29.8740C12:0
7.Methyl tridecanoate32.4720C13:0
8.Methyl myristate34.9740C14:0
9.Methyl myristoleate37.0120C14:1 (c9)
10.Methyl pentadecanoate37.3720C15:0
11.Methyl pentadecenoate39.3620C15:1 (c10)
12.Methyl palmitate39.6860C16:0
13.Methyl palmitoleate41.3020C16:1 (c9)
14.Methyl heptadecanoate41.8620C17:0
15.Methyl heptadecenoate43.4520C17:1 (c10)
16.Methyl stearate43.9740C18:0
17.Methyl octadecenoate44.9220C18:1 (t9)
18.Methyl oleate45.3440C18:1 (c9)
19.Methyl linolelaidate46.3920C18:2 (t9,t12)
20.Methyl linoleate47.3220C18:2 (c9,c12)
21.Methyl arachidate47.9240C20:0
22.Methyl linolenate48.7720C18:3 (c6,c9,c12)
23.Methyl eicosenoate49.2020C20:1 (c11)
24.Methyl linolenate 49.5420C18:3 (c9,c12,c15)
25.Methyl heneicosanoate49.7820C21:0
26.Methyl eicosadienoate51.0920C20:2 (c11,c14)
27.Methyl behenate51.6040C22:0
28.Methyl eicosatrienoate 52.5220C20:3 (c8,c11,c14)
29.Methyl erucate52.8820C22:1 (c13)
30.Methyl eicosatrienoate 53.2820C20:3 (c11,c14,c17)
31.Methyl arachidonate53.4220C20:4 (c5,c8,c11,c14)
32.Methyl tricosanoate53.6520C23:0
33.Methyl docosadienoate54.8520C22:2 (c13,c16)
34.Methyl lignocerate55.3240C24:0
35.Methyl eicosapentaenoate56.0920C20:5 (c5,c8,c11,c14,c17)
36.Methyl nervonate56.7420C24:1 (C15)
37.Methyl docosahexaenoate62.1720C22:6 (c4,c7,c10,c13,c16,c19)
image
GC_FF1261
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
SampleFood industry FAME mix (cat.# 35077)
Diluent:Hexane/dichloromethane
Conc.:1,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:225 °C
Oven
Oven Temp.:100 °C (hold 4 min) to 240 °C at 3 °C/min (hold 15 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorFID @ 285 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:30 mL/min
Air flow:300 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Fats and Oils

,

Nutritional Labeling

,

Trans Fats

,

AOAC 996.06

,

Fatty Acid Profiling

,

High Resolution GLC

,

FAMEs

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us