share:
delim
delim
trans Fat Reference Standard on Rt-2560

trans Fat Reference Standard on Rt-2560

PeakstR (min)Conc.
(wt.%)
Structural Nomenclature
1.Methyl myristelaidate21.8066C14:1 (trans-9)
2.Methyl trans-pentandecen-10-oate24.8326C15:1 (trans-10)
3.Methyl palmitelaidate27.6056C16:1 (trans-9)
4.Methyl trans-heptadecen-10-oate30.6606C17:1 (trans-10)
5.Methyl petroselaidate33.4074C18:1 (trans-6)
6.Methyl elaidate33.51316C18:1 (trans-9)
7.Methyl trans-vaccenate33.67912C18:1 (trans-11)
8.Methyl linoelaidate35.5836C18:2 (trans-9,12)
9.Methyl trans-nonadecen-7-oate36.3036C19:1 (trans-7)
10.Methyl trans-nonadecen-10-oate36.45512C19:1 (trans-10)
11.Methyl trans-eicosan-11-oate39.2706C20:1 (trans-11)
12.Methyl brassidate44.7646C21:1 (trans-13)
13.Methyl ricinelaidate56.9878C18:1 (trans-9; OH-12)
image
GC_FF1267
ColumnRt-2560, 100 m, 0.25 mm ID, 0.20 µm (cat.# 13198)
Sampletrans Fat reference standard (cat.# 35629)
Diluent:Hexane
Injection
Inj. Vol.:1 µL split (split ratio 200:1)
Liner:Topaz 4.0 mm ID straight inlet liner w/wool (cat.# 23300)
Inj. Temp.:250 °C
Oven
Oven Temp.:120 °C (hold 1 min) to 250 °C at 2 °C/min (hold 4 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
InstrumentAgilent 7890A GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us